Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
References
Configurations
No configuration.
History
04 Jun 2026, 22:16
| Type | Values Removed | Values Added |
|---|---|---|
| Summary | (en) Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted. |
26 May 2026, 18:16
| Type | Values Removed | Values Added |
|---|---|---|
| New CVE |
Information
Published : 2026-05-26 18:16
Updated : 2026-06-04 22:16
NVD link : CVE-2026-7251
Mitre link : CVE-2026-7251
CVE.ORG link : CVE-2026-7251
JSON object : View
Products Affected
No product.
CWE
CWE-259
Use of Hard-coded Password