CVE-2026-0611

Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
Configurations

No configuration.

History

02 Jun 2026, 20:16

Type Values Removed Values Added
References () https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf - () https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf -

02 Jun 2026, 17:16

Type Values Removed Values Added
New CVE

Information

Published : 2026-06-02 17:16

Updated : 2026-06-02 20:16


NVD link : CVE-2026-0611

Mitre link : CVE-2026-0611

CVE.ORG link : CVE-2026-0611


JSON object : View

Products Affected

No product.

CWE
CWE-306

Missing Authentication for Critical Function